dc.identifier.citationContreras Vélez, J. J. (2020). Revisión de métodos de calibración para monitores fetales. Universidad Santiago de Cali.es
dc.identifier.urihttps://repository.usc.edu.co/handle/20.500.12421/4253-
dc.descriptionLa normatividad nacional vigente exige a las Instituciones Prestadoras de Servicios de Salud calibrar los equipos biomédicos que contengan variables medibles y que sean empleados en el ámbito hospitalario. Dicha actividad genera mayor confiabilidad en el funcionamiento de los equipos e incrementa la calidad en los servicios de atención en salud. Para realizar la calibración, es necesario crear y validar métodos para cada equipo teniendo en cuenta las variables que hacen parte de su sistema. Actualmente, la superintendencia de industria y comercio (SIC), quien se encarga de emitir reglamentos técnicos para la estandarización de métodos, no ha emitido dichos lineamientos para todos los equipos biomédicos, por lo tanto, cada empresa que preste el servicio de calibración debe crear y validar sus propios métodos (métodos no estandarizados). Uno de los equipos que no cuenta con métodos estandarizados son los monitores fetales, los cuales son equipos de diagnóstico empleados por el personal médico para evaluar información relacionada a la frecuencia cardiaca fetal y actividad uterina. Teniendo en cuenta su importancia en la práctica clínica, primero se identifica el principio de funcionamiento, las variables que mide y se analizan los rangos de operación pertinentes para cada variable en el ámbito clínico. Luego se explica brevemente el uso y la información que se evalúa en un examen cardiotocográfico (CTG). Por último, se realiza una revisión de los métodos de calibración existentes para monitores fetales. Es importante resaltar que los métodos mencionados para cada variable, se toman de diversos autores enfatizando en aquellos en los que se encuentra información certificada por entes acreditadores, donde se evalúa el método y el instrumento de medición o simulador aplicadoes
dc.description.abstractCurrent national regulations requires health service providers to calibrate biomedical equipment containing measurable variables and used in hospitals. This activity generates greater reliability in the operation of the equipment and increases the quality of health care services. To perform the calibration, it is necessary to create and validate methods for each piece of equipment, taking into account the variables that are part of its system. Currently, the Superintendence of Industry and Commerce (SIC), which is responsible for issuing technical regulations for the standardization of methods, has not issued such guidelines for all biomedical equipment; therefore, each company providing calibration services must create and validate its own methods (non-standard methods). One of the equipment that does not have standardized methods is fetal monitors, which are diagnostic equipment used by medical personnel to evaluate information related to fetal heart rate and uterine activity. Considering their importance in clinical practice, the principle of operation is identified, the variables it measures and the operating ranges relevant to each variable in the clinical setting are analyzed. The use and information evaluated in a cardiotocography test (CTG) is then briefly explained. Finally, a review of existing calibration methods for fetal monitors is provided. It is important to note that the methods mentioned for each variable are taken from various authors with emphasis on those where information is certified by accrediting bodies, where the method and the measuring instrument or simulator applied is evaluated.es
dc.formatapplication/pdf-
dc.format.extent25 páginas-
dc.language.isoeses
dc.publisherUniversidad Santiago de Calies
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/-
dc.subjectMétodo de Calibraciónes
dc.subjectMonitor Fetales
dc.subjectDoppleres
dc.subjectCardiotocografíaes
dc.subjectEquipo Patrónes
dc.subjectIncertidumbrees
dc.titleRevisión de métodos de calibración para monitores fetaleses
dc.typeThesises
dc.publisher.facultyFacultad de Ingeniería-
dc.pubplace.cityCali-
dc.rights.accesoAcceso Abierto-
dc.rights.ccReconocimiento 4.0 Internacional (CC BY 4.0)-
dc.source.repositoryRepositorio Institucional USC-
dc.source.institutionUniversidad Santiago de Cali-
dc.type.spaArtículo-
dc.creator.degreeTrabajo de grado para optar al título de Bioingeniería-
dc.publisher.programBioingeniería-
dc.pubplace.stateValle del Cauca-
dc.subject.keywordCalibration Method-
dc.subject.keywordFetal Monitor-
dc.subject.keywordDoppler-
dc.subject.keywordCardiotocography-
dc.subject.keywordStandard Equipment-
dc.subject.keywordUncertainty-